Molecular Sieve

Molecular Sieve Trap

These traps perform a dual function. A zeolite desiccant removes hydrocarbons which backstream toward the chamber when the mechanical pump oil is at it’s vapor pressure. Also, the zeolite traps water vapor and other gasses before they reach the mechanical pump and contaminate the pump oil. Depending on the process, this can reduce the oil change frequency significantly.

Ebara's molecular sieve trap utilizes the two-piece trap design for ease of maintenance and zeolite replacement. For maximum conductance the sieve material is contained in a suspended basket fabricated from perforated 304 stainless steel. This design provides conductance to the sieve at the top of the basket as well as around the periphery. A mid-basket baffle causes the gas to enter one-half of the basket and to exit the other half.

A molecular sieve charge and regeneration heater are included with every trap. 240V regeneration heater is available as an option. Other voltages and wattages are available upon request. The heater is held in place with a clip that allows trap mounting in any position. Standard finish is beadblasted. Electropolish finish, other body and flange configurations, and larger body sizes are available.

Replacing The Sieve

When initializing a trap, replacing the sieve or when base pressure cannot be achieved, the sieve should be regenerated by turning on the heater for a few hours and running the mechanical pump with it's ballast valve open. The frequency and duration of sieve regeneration is dependent upon the kind and amount of gasses produced by the particular application. To replace the sieve, remove the trap end that contains the heater and remove the nut at the end of the basket. Dump the expired sieve material, replace with new, reattach the basket/end plate then perform the regeneration process outlined above.

Specifications

Choose product to view specifications

In-Line

Material
   Trap Body: 304 stainless steel
   Flanges: 304 stainless steel
   Desicant: Zeolite 13X, 1/8? pellets
Vacuum range: >= 1 x 10-8 Torr
Temperature range: -20°C to 150°C
Options: 240V regeneration heaters and ISO body flanges

Additional Information

Model Info

MODEL NUMBER BODY
OD
FLANGE
TYPE
PORT
OD
REGENERATION
HEATER RATING
A B C
OILTM-4-0752-KFB 4 KF-16B 3/4 120V/125W 6.20 7.30 3.50
OILTM-4-1002-KFB 4 KF-25B 1 120V/125W 6.20 7.30 3.50
OILTM-4-1502-KFB 4 KF-40B 1-1/2 120V/125W 6.20 7.30 3.50
OILTM-6-1502-KFB 6 KF-40B 1-1/2 120V/230W 8.10 9.30 4.00
OILTM-6-2002-KFB 6 KF-50B 2 120V/230W 8.10 9.30 4.00
OILTM-8-2002-KFB 8 KF-50B 2 120V/350W 13.00 11.30 6.75
OILTM-8-2502-ISO 8 ISO-63-250-OF 2-1/2 120V/350W 13.00 11.50 6.75
OILTM-8-3002-ISO 8 ISO-80-300-OF 3 120V/350W 13.00 11.50 6.75
OILTM-8-4002-ISO 8 ISO-100-400-OF 4 120V/350W 13.00 11.50 6.75

Angle

Material
   Trap Body: 304 stainless steel
   Flanges: 304 stainless steel
   Desicant: Zeolite 13X, 1/8? pellets
Vacuum range: >= 1 x 10-8 Torr
Temperature range: -20°C to 150°C
Options: 240V regeneration heaters and ISO body flanges

Additional Information

Model Info

MODEL NUMBER BODY
OD
FLANGE
TYPE
PORT
OD
REGENERATION
HEATER RATING
A B C
ATM-4-0752-KFB 4 KF-16B 3/4 120V/125W 7.70 6.55 3.65
ATM-4-1002-KFB 4 KF-25B 1 120V/125W 7.70 6.55 3.65
ATM-4-1502-KFB 4 KF-40B 11/2 120V/125W 7.70 6.55 3.65
ATM-6-1502-KFB 6 KF-40B 11/2 120V/230W 9.60 7.75 4.65
ATM-6-2002-KFB 6 KF-50B 2 120V/230W 9.60 7.75 4.65
ATM-8-2002-KFB 8 KF-50B 2 120V/350W 14.65 11.53 5.65
ATM-8-2502-ISO 8 ISO-63-250-OF 21/2 120V/350W 14.75 11.63 5.75
ATM-8-3002-ISO 8 ISO-80-300-OF 3 120V/350W 14.75 11.63 5.75
ATM-8-4002-ISO 8 ISO-100-400-OF 4 120V/350W 14.75 11.63 5.75

Straight In-Line

Material
   Trap Body: 304 stainless steel
   Flanges: 304 stainless steel
   Desicant: Zeolite 13X, 1/8" pellets
Vacuum range: >= 1 x 10-8 Torr
Temperature range: -20°C to 150°C
Options: 240V regeneration heaters and ISO body flanges

Additional Information

Model Info

MODEL NUMBER BODY
OD
FLANGE
TYPE
PORT
OD
REGENERATION
HEATER RATING
A B C
ILTM-4-0752-KFB 4 KF-16B 3/4 120V/125W 7.00 7.30 3.50
ILTM-4-1002-KFB 4 KF-25B 1 120V/125W 7.00 7.30 3.50
ILTM-4-1502-KFB 4 KF-40B 11/2 120V/125W 7.00 7.30 3.50
ILTM-6-1502-KFB 6 KF-40B 11/2 120V/230W 7.00 9.30 3.50
ILTM-6-2002-KFB 6 KF-50B 2 120V/230W 7.00 9.30 3.50
ILTM-8-2002-KFB 8 KF-50B 2 120V/350W 10.00 11.30 5.00
ILTM-8-2502-ISO 8 ISO-63-250-OF 21/2 120V/350W 10.00 11.50 5.00
ILTM-8-3002-ISO 8 ISO-80-300-OF 3 120V/350W 10.00 11.50 5.00
ILTM-8-4002-ISO 8 ISO-100-400-OF 4 120V/350W 10.00 11.50 5.00